C-2 72A-1230 Rev. F
July 22, 2005
M4000 Firmware/Application Errors
31 Unused
32 Internal Amp Cable Disconnected
33 One or Both Safety Switch Cables Disconnected
34 Power Amp Power Bad
35 Overcurrent on Power Amp Output
36 Power Amp DC Supply High
37 Power Amp DC Supply Low
38 M4000 Synchronization Error
39 Phase Lock Loop Not Locked
40 Strobe Cable Not Plugged In
41 Strobe Not Flashing
42 - 49 Reserved for Future Use
50 Lead Changed During Test
51 Safety Switches Not Released Between Tests
52 Software Buffer Overrun
53 One or Both Low Voltage Leads Disconnected
54 Overcurrent Occurred
55 Amplifier OverTemperature Occurred
56 Input Current Peak Maximum Exceeded
57 Transformer OverTemperature Occurred
58 Test Stopped By Controller
59 Comm. Timeout Detected By Instrument
60 Main Reference OverTemperature Occurred
61 Calibration Reference OverTemperature Occurred
62 High Voltage Over 12.5 kV
63 Voltage Input Saturated
64 Requested Maximum Voltage Exceeded
Error # Description of Error
Error # Description Of Error
81 Unable to open application file