Test Results
3-8 72A-1230 Rev. F
July 22, 2005
Figure 3.5 Test in Progress
The test will proceed automatically if Auto Ramp Mode is selected in the
Clipboard Configuration, or will be ramped up by the operator if Manual Set
Voltage is selected in the Clipboard Configuration. The digital values for these
parameters will be shown to the right of the meters. The table on the screen
describes test conditions and power factor test results. The line on the bottom
of the screen displays the status of the test. When running multiple Clipboard
tests, the line number of the test that is running will be displayed on the same
line. For each test line, the voltage will ramp to the required maximum test
voltage twice. This is required to implement the selected electrostatic
interference cancellation technique (Line Synchronized Reversal or Line
Frequency Modulation).
Test Results
After the test is completed, the following appears: