M4000 Insulation Analyzer User Guide
72A-1230 Rev. F 3-5
July 22, 2005
Corr. Fact. Temperature Correction Factor. This column includes
automatic application of temperature correction factors
to the Clipboard Form. The default correction factor is
1.00.
The Measured Percent Power Factor is multiplied by this
Temperature Correction Factor to obtain the Corrected
Percent Power Factor in the column to the right of the
Corr. Fact. Column. See ”Temperature Correction”
on page 2-12 for this automatic function.
Corr%PF Corrected Percent Power Factor. In this column, you will
find the results of the Measured Percent Power Factor
Times the Correction Factor. You cannot enter or change
results in this column.
Cap. /Ind. Capacitance/Inductance. Whether the current is
capacitive or inductive is displayed in this column. You
cannot enter or change results in this column.
RT Rating. You may enter a one letter rating (G. B. I., etc.)
of the insulation in this column.
N Note. If a note is entered for a single line, a check mark
will show in this box. A test note is entered by selecting
Test Note... from the Tools pull-down menu item, or by
clicking on the notepad icon. You will be asked to which
test line the note corresponds. The default test line is No.
1.
To enter a note that relates to the entire Clipboard Form,
type in the area at the bottom of the screen titled, Sheet
Note (If it is hidden, click Sheet Note from the View
menu).
To clear a note, select Clear from the Tools pull-down
menu item.
Column Description