Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 141
Alarms
Viewing a snapshot of a single event
20 seconds of waveform and parameter data are
stored automatically in the alarm history under the
following circumstances:
A parameter whose recording archive feature is
set to Store or Str/Rec (see page 126) violates
set alarm limits.
You select the Mark event button on the main
menu bar.
In both instances, events with stored waveform and
parameter data are identified on the alarm history
by the following symbol . Such an event
consists of a snapshot of all connected parameter
values and waveforms. Of the 20-second event
capture, 10 seconds were recorded before and
10 seconds were recorded after the event
occurred.
To view a snapshot of a stored event
1 Select Alarms... on the main menu bar.
2 Select the Alarm history tab.
3 Select the row of the event marked with the
symbol that you wish to view.
The following diagram shows the event snapshot
screen.
A Event header showing the date, time, duration,
priority, and alarm message
B Parameter values area
C Print button for printing the alarm event report
D Zoom out button
E Zoom in button
F Navigation arrows for scrolling through events
G Select Event button
H Waveform area
NOTE
To return to the alarm history, select the Select
Event button.
022
A
B
CD
E
F
G
F
Alarms
H

Table of Contents