EasyManuals Logo

Dräger Infinity C700 User Manual

Dräger Infinity C700
610 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #160 background imageLoading...
Page #160 background image
Alarms
160 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Arrhythmia ranges and defaults
Parameter Alarm priority
default
Rate (default) Count (default) Alarm archive
factory default
ASY High Not adjustable Not adjustable Str/Rec
VF High Not adjustable Not adjustable Str/Rec
VTACH High 100 to 200 (120)
Increments of 10
5 to 15 (10)
Increments of 1
Str/Rec
ARTF Off Not adjustable Not adjustable Off
RUN Medium not adjustable (Rate =
VTACH)
3 to VT count – 1 (3 to
9) changes based on
VTACH
Str/Rec
AIVR Medium Not adjustable =
VTACH rate – 1 (119)
Not adjustable (3) Off
SVT Medium 120 to 200 (150)
Increments of 10
3 to 10 (3)
Increments of 1
Str/Rec
CPT Low Not adjustable Not adjustable Str/Rec
BGM Low Not adjustable Not adjustable Str/Rec
TACH Off 100 to 200 (130)
Increments of 10
5 to 15 (8)
Increments of 1
Off
BRADY Off 30 to 105
(adult 50; pediatric
60)
Increment of 5
Not adjustable
(8)
Off
Pause Off 1 to 3.5 (2.5)
Increments of 0.5
Not adjustable Off

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity C700 and is the answer not in the manual?

Dräger Infinity C700 Specifications

General IconGeneral
BrandDräger
ModelInfinity C700
CategoryMedical Equipment
LanguageEnglish

Related product manuals