Alarms
160 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Arrhythmia ranges and defaults
Parameter Alarm priority
default
Rate (default) Count (default) Alarm archive
factory default
ASY High Not adjustable Not adjustable Str/Rec
VF High Not adjustable Not adjustable Str/Rec
VTACH High ≥100 to 200 (≥120)
Increments of 10
≥5 to 15 (≥10)
Increments of 1
Str/Rec
ARTF Off Not adjustable Not adjustable Off
RUN Medium not adjustable (Rate =
VTACH)
3 to VT count – 1 (3 to
9) changes based on
VTACH
Str/Rec
AIVR Medium Not adjustable =
VTACH rate – 1 (≤119)
Not adjustable (≥3) Off
SVT Medium ≥120 to 200 (≥150)
Increments of 10
≥3 to 10 (≥3)
Increments of 1
Str/Rec
CPT Low Not adjustable Not adjustable Str/Rec
BGM Low Not adjustable Not adjustable Str/Rec
TACH Off ≥100 to 200 (≥130)
Increments of 10
≥5 to 15 (≥8)
Increments of 1
Off
BRADY Off ≤30 to 105
(adult ≤ 50; pediatric
≤60)
Increment of 5
Not adjustable
(≥8)
Off
Pause Off 1 to 3.5 (2.5)
Increments of 0.5
Not adjustable Off