Cardiac output (C.O.)
334 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Accessing the cardiac output settings
Select the cardiac output parameter field to
select the C.O. page directly.
or
1 Select Sensor parameters... from the main
menu bar.
2 Select the C.O. tab to access the C.O. page. If
you cannot see the tab, select the following two
symbols located in the upper right corner of the
dialog: symbol and the display filter
button.
The top portion of the page contains the Auto set
and Alarm buttons for configuring the alarm
functions. For detailed alarm setup information, see
"Configuring the alarm settings for a patient" on
page 125.
Cardiac output parameter setup functions
All cardiac output setup functions take place in the
C.O. page.
Selection Available settings Description
Catheter type – BD/Ohmeda (default)
– Edwards/Baxter
– Arrow
–Other
Displays the currently selected catheter
type.
Catheter size 5F, 7F (default), 7.5F Displays the currently selected catheter
size.
If Other is selected for Catheter type
setting, this button is not available.
Injectate volume
[cc]
3.0, 5.0, or 10.0 (default) Displays the currently selected volume of
the injectate.
If Other is selected for Catheter type
setting, this button is not available.