Reports/recordings
496 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Automatic alarm recordings
When the Alarm Waveform feature is activated
(see page 467), timed alarm recordings are
generated automatically whenever a parameter
whose archive function is activated goes beyond
the set alarm limits.
Alarm recordings are also generated when an
arrhythmia event with an alarm classification of
high or medium occurs.
The following sections describe how to set up a
parameter or arrhythmia event to generate an
automatic alarm recording.
To activate or deactivate the archive function of
a parameter
1 Select the parameter field of the parameter
whose alarm function you wish to activate or
deactivate to access that parameter page
directly.
or
Select Sensor parameters... from the main
menu bar > select the desired parameter tab to
access the page.
2 Select the button next to the Archive setting
and select either Store, Str/Rec, Record to
generate a recording or Off to deactivate the
feature.
To assign an alarm priority to arrhythmia events
1 Select the Alarms... button on the main menu
bar.
2 Select the Limits tab (if not already selected).
3 Select the ARR tab along the right side to
display the ARR page.
4 Touch the field in the Alarm column of the
parameter whose alarm priority you want to
modify.
5 Select the field in the Archive column and
select either Store, Str/Rec, Record to
generate a recording or Off to deactivate the
feature.
6 Press the rotary knob to confirm the setting.