EasyManua.ls Logo

Dräger Infinity C700 - CO2 Display; CO2 Parameter Field

Dräger Infinity C700
610 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Scio Monitoring
366 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
CO2 display
On the Cockpit, the CO2 display consists of:
–CO
2 parameter field
–CO
2 waveform
The Cockpit supports a configurable parameter
field for display of CO
2 that allows one of the
following displays:
–CO2 by itself (default)
–CO
2 with O2 (only for gas analyzers with O2
monitoring)
CO2 parameter field
The default CO2 parameter field displays the
current values for:
Inspired CO
2 (inCO2) – the level of CO2 in the
airway during inspiration phase.
End-tidal CO
2 (etCO2) – the level of CO2 in the
airway at the end of expiration.
Respiratory Rate (RRc) – the patient’s
respiratory rate, derived from the etCO2 signal.
A etCO2 label
B etCO
2 value – highest level of CO2 in the airway
during the expiratory phase within the
measurement interval
C etCO2 alarm limits (displays crossed triangle
symbol when etCO2 alarms are turned off)
D inCO
2 label
E inCO
2 value – the level of CO2 in the airway
during inspiration, taken as the minimum value
within the measurement interval
F inCO2 alarm upper limit (displays crossed
triangle symbol when the inCO2 alarm is turned
off)
G RRc alarm limits (displays crossed triangle
symbol when RRc alarms are turned off)
H RRc parameter label
I RRc value – respiratory rate derived from the
CO
2 signal.
EtcO2pbox_en.pdf
27
0
21
etCO2
inCO2
RRc
70
20
30
5
4
A
B
CDEF
GHI

Table of Contents