Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 517
Troubleshooting
ECG
Priority Message Parameter field Problem Solution
!!! Asystole ASY The reported
arrhythmia was
detected
Check the patient and
treat if necessary.
!!! Bradycardia
(neonatal patient
category)
BRADY The reported
arrhythmia was
detected
Check the patient and
treat if necessary.
!! HR > (alarm limit)
HR < (alarm limit)
Parameter value The parameter
value is
above/below the
set upper/lower
alarm limits.
Check the patient
and treat if
necessary.
Change the alarm
limits.
!! %0 out of range
high
1)
The parameter value
is replaced by +++
The parameter
value is above the
measurement
range of the
monitor.
! %0 artifact
1) 2)
The parameter value
is replaced by ***
Patient
movement
(shivering,
tremors)
Bad electrode
contact
Excessive
signal noise
interference
from auxiliary
equipment
Check the
electrodes and
reapply if
necessary.
Make sure that
the patient’s skin
is properly
prepped.
Isolate the patient
from auxiliary
equipment, if
possible.
1)
%0 is a placeholder for the parameter label HR or ECG.
2)
After discharging a patient or starting the device, the alarm becomes active only after a numeric value
has been received for that parameter.