Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 367
Scio Monitoring
CO2/O2 parameter field
The CO2 /O2 parameter field displays the following
additional values:
–Inspired O
2 (in O2/FiO2) – the level of O2 in the
airway during inspiration, taken as the minimum
value within the measurement interval
– End-tidal O2 (etO2) – the highest level of O2 in
the airway during the expiratory phase within
the measurement interval
A CO
2 label
B O
2 label
C Abbreviation for “inspiratory”
D inCO
2 value – the level of CO2 in the airway
during inspiration, taken as the minimum value
within the measurement interval
E inO2 (FiO2) value – the level of O2 in the airway
during inspiration, taken as the minimum value
within the measurement interval
F inCO
2 alarm upper limit (displays crossed
triangle symbol when the inCO
2 alarm is turned
off)
G inO2 (FiO2) alarm limits (displays crossed
triangle symbol when inO
2 alarms are turned
off)
H Abbreviation for " expiratory"
I etCO
2 value – highest level of CO2 in the airway
during the expiratory phase within the
measurement interval
J etO2 value – highest level of O2 in the airway
during the expiratory phase within the
measurement interval
K etCO2 alarm limits (displays crossed triangle
symbol when etCO2 alarms are turned off)
L etO
2 alarm limits (displays crossed triangle
symbol when etO2 alarms are turned off)
CO2 waveform
The M540 also displays an instantaneous CO2
waveform.
A Expiratory or alveolar plateau (level of CO
2 in
lungs ceases to increase significantly)
B End-tidal concentration point (end of expiratory
phase, where CO2 is measured)
C Onset of inspiratory phase
D Onset of expiratory phase
E Baseline during inspiration
NOTE
Current RRc values are displayed on the CO2 tab
of the Scio dialog.