Microstream CO2 monitoring
354 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
CO2 parameter setup
Setup functions for CO2 parameters take place in
the CO2 dialog within the Microstream tab.
For detailed alarm setup information, see "Config-
uring the alarm settings for a patient" on page 125.
NOTE
When a Scio module is connected, parameter
controls for CO
2 are available only in the Scio
setup menu.
Selection Available settings Description
RRc apnea time
[s]
Off (default)
10 s, 15 s, 20 s, 25 s, 30 s
Specifies how long the Cockpit waits before
reporting a cessation in breathing as an apnea
event.
RRc apnea
archive
Off, Store (default), Str/Rec Determines what happens in response to an
apnea event.
Next service in: Informational only (settings
are not applicable)
The remaining hours until maintenance is
required.
Averaging Last valid breath,
Off,
10 s, 20 s (default), 30 s
Controls the specific time or the interval used
to select the maximum measured etCO
2 and
the minimum measured inCO2.
Scale – 0 to 40 mmHg (default), 0
to 60 mmHg,
0 to 80 mmHg, 0 to
100 mmHg
– 0 to 5 kPa (default), 0 to
8 kPa, 0 to 12 kPa, 0 to
16 kPa
– 0 to 5% (default),
0 to 8%, 0 to 12 %,
0 to 16%
Adjusts the size of the CO
2 waveform.
Scale settings apply to both Mainstream and
Microstream.
Color Red, White, Yellow (default),
Green, Light blue, Blue,
Purple, Orange
Determines the color of the CO
2 waveform,
and the parameter labels and values.
Last calibration: Informational only Displays the date of the last calibration.