Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 593
Technical data
Sound pressure
Secondary display
A secondary display has to meet the minimum
technical specifications outlined in the following
table.
Starting with IACS VG5.0, the maximum resolution
of the secondary display must meet the
requirements provided in the table. For information
on how to connect a secondary display to the IACS,
see page 58.
Electromagnetic compatibility
The separation distances are written with regard to
the Cockpit. The numbers provided will not
guarantee faultless operation but should provide
reasonable assurance of such. This information
may not be applicable to other medical electrical
equipment, and older equipment may be
particularly susceptible to interference.
Sound pressure levels for IEC alarm tones
Measurements per ISO 3744, 5 to 100% volume setting
Device Low priority Medium priority High priority
C500 (1
st
generation) 50 dB(A) to 68 dB(A) 53 dB(A) to 69 dB(A) 55 dB(A) to 71 dB(A)
C700 (1
st
generation) 53 dB(A) to 70 dB(A) 54 dB(A) to 71 dB(A) 55 dB(A) to 73 dB(A)
C500 (2
nd
generation) 50 dB(A) to 68 dB(A) 55 dB(A) to 73 dB(A) 56 dB(A) to 74 dB(A)
C700 (2
nd
generation) 48 dB(A) to 69 dB(A) 52 dB(A) to 73 dB(A) 53 dB(A) to 74 dB(A)
General requirements
Resolution 50.8 cm (20 in) display: 1680 x 1050
43.2 cm (17 in) display: 1440 x 900
Maximum supported distance 5 m (16.4 ft)
Aspect ratio 16:10 (1
st
and 2
nd
generation)
Display delay 250 ms in reference to the patient signal
Connection to Cockpit DVI-I 1 connector only
Standards compliance IEC60950