Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 307
Invasive blood pressure (IBP)
Connecting the Dual Hemo MCable
The Dual Hemo MCable connects directly to the
M540.
A Hemodynamic port on the M540
B Dual Hemo MCable connector
C Transducer adapter cable
D Transducer
E Transducer cables
To connect the Dual Hemo MCable
1 Attach the transducers (D) to the transducer
adapter cables (C). The transducer adapter
cables are permanently fastened to the Dual
Hemo MCable.
2 Connect the Dual Hemo MCable connector (B)
to the gray hemodynamic port (A) on the M540.
Preventing fluid ingress
Refer to the following figure to correctly position the
Dual Hemo MCable to prevent fluids from entering
the ports where the transducer cables are attached.
344
A
B
E
D
C
C