External devices – Servo-i ventilator
422 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Supported Servo-i parameters
The following table lists which Servo-i ventilator pa-
rameters are displayed and trended on the Cockpit.
Refer to page 421 for a list of supported ventilators
and the compatible software versions.
Parameter
label
Description Unit of
mea-
sure-
ment
Displayed in Ventila-
tion Show all page
(see page 426)
Yes/No?
Cockpit
Trends page
Available
for display
in parame-
ter field
Yes/No?
Cdyn Dynamic compliance L/bar Yes Continuous No
etCO2 End-tidal CO2
concentration
mmHg Not supported for
display
Continuous and
mini-trend
Yes
kPa
%
FiO2 Inspired O2 % Yes Continuous No
MVe Minute volume,
expired
L/min Yes Continuous and
mini-trend
Yes
MVespon Spontaneous
expiratory minute
volume
L/min Yes Continuous No
PEEP Peak end expiratory
airway pressure
cmH2O Yes Continuous Yes
PIP Peak inspiratory
pressure
cmH2O Yes Continuous and
mini-trend
Yes
Pmean Mean airway
pressure
cmH2O Yes Continuous Yes
RR Respiratory rate /min Yes Continuous Yes
V'CO
2 Carbon dioxide
production
mL/min Yes Continuous No
VTe Tidal volume,
expired
mL Yes Continuous Yes