EasyManua.ls Logo

Dräger Infinity C700 - BIS Parameter Setup Functions; BIS Show All Page; Accessing the bis Settings

Dräger Infinity C700
610 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
External Device – Bispectral index (BIS)
384 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Accessing the BIS settings
Select the BIS parameter field to select the BIS
page directly.
or
1 Select Sensor parameters... from the main
menu bar.
2 Select the BIS tab to access the BIS page.
If you cannot see the tab, select the following
two symbols located in the upper right corner of
the dialog: symbol and the display filter
button.
3 Select the Settings and Show all tab.
BIS parameter setup functions
All BIS setup functions take place in the BIS page.
Selection Available settings Description
Settings page
Scale[uV] 5, 10, 25, 50, 100 (default), 250 µV Selects the scale of the EEG T
waveform.
Smoothing rate
[s]
Informational data only Not applicable
BIS secondary
parameter
BSR, BCT (default), SEF, PWR Selects a secondary parameter
and assigns it to the BIS
parameter field (see page 381).
No alarm
signaling for this
device.
Informational message that no optical or
acoustic alarm signals are available on the
the Cockpit
Not applicable
BIS Show all page
This page displays all supported BIS parameters, settings, labels and the units of measure where
appropriate.

Table of Contents

Related product manuals