Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 537
Troubleshooting
Temperature
Priority Message Parameter field Problem Solution
!! T1a > (alarm limit)
T1b > (alarm limit)
Ta > (alarm limit)
Tb> (alarm limit)
T > (alarm limit)
T > (alarm limit)
T1 > (alarm limit)
T1a < (alarm limit)
T1b < (alarm limit)
Ta < (alarm limit)
Tb < (alarm limit))
T < (alarm limit)
T < (alarm limit)
T1 < (alarm limit)
Parameter value The parameter
value is
above/below the
upper/lower
alarm limits.
Check the
patient and
treat if
necessary.
Change the
alarm limits.
!! %0 out of range
high
1)
The parameter value
is replaced by +++
The parameter
value is
above/below the
measurement
range of the
monitor.
Check the
patient and
treat if
necessary.
Check the
equipment and
replace, if
necessary.
!! %0 out of range low
1)
The parameter value
is replaced
by - - -
! Cannot derive %0
2)
The parameter value
is replaced by ***
The cable is
either faulty or
unplugged.
Check the
equipment and
replace it if
necessary.
Connect the
second
temperature
sensor.
! %0 H/W failure
1), 2)
The parameter value
is replaced by ***
The hardware
reference values
do not meet the
specified
tolerance.
Contact
DrägerService
®
.
! %0 Unplugged
1), 2)
The parameter value
is replaced by ***
The temperature
sensor is
unplugged.
Reapply the
temperature
sensor.