External Device – Neuromuscular transmission (NMT)
392 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
NMT parameter functions and settings
Selection Available settings Description
Display
temperature
–On
– Off (Default)
Controls the temperature
display in the NMT parameter
field.
Stim current 0 to 60 mA (no default) Displays the stimulation current
of the NMT device.
Stim current
mode
Auto The NMT device establishes a
supramaximal current during
the first measurement and uses
it for subsequent
measurements.
Manual The stimulation current was
selected manually.
Sensitivity (1-512) Between 1 and 512 (no default) Displays the sensitivity of the
NMT device.
Pulse width 200 or 300 µs Determines the width of the
pulse.
Print Prints the current NMT settings and up to 500 of the most recent measurements.
No alarm
signaling for this
device.
Informational message that no optical or acoustic alarm signals are available on
the Cockpit.