Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 539
Troubleshooting
Invasive blood pressure (IBP)
Priority Message Parameter field Problem Solution
!!! ART cath.
disconnected?
The parameter value
is replaced by ***
The arterial
catheter could
be dislodged, or
there could be a
leak in the
tubing.
Assess the catheter
insertion site.
Inspect the tubing for
leaks or the
presence of blood.
Check the patient
and treat, if
necessary.
!! %0 transducer
failure
1)
Pressure
transducer
hardware
failure.
Check the
transducer and
replace, if necessary.
!! IBP label > (alarm
limit)
IBP label < (alarm
limit)
Parameter value The parameter
value is
above/below the
upper/lower
alarm limits.
Check the patient
and treat if
necessary.
Change the alarm
limits.
!! CPP > (alarm limit)
CPP < (alarm limit)
Parameter value The parameter
value is
above/below the
upper/lower
alarm limits.
Check the patient
and treat if
necessary.
Change the alarm
limits.
!! %0 out of range
low
1)
The parameter value
is replaced by - - -
The parameter
falls outside the
pressure range
of the monitor.
Check the patient
and treat if
necessary.
Check the
equipment and
replace, if necessary.
!! %0 out of range
high
1)
The parameter value
is replaced by +++
! %0 please check
zero
1)
Parameter value The IBP zero
value stored in
the M540 was
lost and the
transducer
requires
zeroing.
Zero the transducer.
1)
%0 is a placeholder for the respective IBP label (including CPP).