Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 305
Invasive blood pressure (IBP)
Connecting the Hemo4 pod and Hemo2 pod
The Hemo4 pod and Hemo2 pod connect directly to
the M540. The following diagram shows where the
gray hemodynamic port (A) is located on the side of
the M540.
A Hemodynamic port on the M540
B Hemodynamic port on the pod
C Pod connection cable
D Transducer cable
E Transducer slot
F Transducer adapter cable
G Invasive blood pressure adapter block
To connect the Hemo4 pod and Hemo2 pod
1 Attach the invasive blood pressure adapter (G)
to the bottom of the Hemo4 pod/Hemo2 pod.
2 Connect one end of the connection cable (C) to
the Hemo4 pod or the Hemo2 pod port (B).
3 Connect the other end of the connection cable
(C) to the gray hemodynamic port of the M540
(A).
4 Attach the transducers to the transducer slot
(E).
5 Connect the transducer adapter cables (F) to
the transducer cable (D).