Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 259
SpO2 and Pulse CO-Ox monitoring with Masimo SET MCable
To apply the sensor
1 Select the size and type of sensor that is best
suited for your patient. Follow the
recommendations of the manufacturer.
2 Position the sensor correctly and attach it to
your patient.
3 Connect the sensor to the Masimo SET MCable
or the Masimo rainbow SET MCable.
SpO2 and Pulse CO-Ox display
On the Cockpit, the SpO2 display consists of:
SpO2 parameter field
A user-configurable Pulse CO-Ox parameter
field when a Masimo rainbow SET MCable with
additional parameters activated is connected.
SpO2 pulse plethysmogram waveform
The following table lists the maximum times the
M540 requires to report the parameter values after
connecting the sensor to the MCable.
330
NOTE
After connecting the sensor, if the sensor-LED
does not light up:
observe the monitor for any message and act
accordingly, or
replace the sensor.
NOTE
The pulse plethysmogram waveform is directly
proportional to the strength of the pulse
amplitude.
Parameter Maximum time
SpO2, PLS, PI Up to 35 s
SpMet, PVI, SpCO Up to 60 s
SpHb, SpOC Up to 90 s
PVI Up to 150 s