370 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
O2 display
On the Cockpit, the O2 display consists of:
–O
2 parameter field
–O
2 waveform
The Cockpit supports a configurable parameter
field for display of O2 that allows one of the
following displays:
–O
2 by itself (default)
–O
2 with N2O
O2 parameter field
The default O2 parameter field displays the current
values for:
– Inspired O
2 (inO2/FiO2) – the level of O2 in the
airway during inspiration phase.
– End-tidal O
2 (etO2) – the level of O2 in the
airway at the end of expiration.
A O
2 label
B Abbreviation for "inspiratory"
C inO2 (FiO2) value – the level of O2 in the airway
during inspiration, taken as the minimum value
within the measurement interval
D inO2 (FiO2) alarm limits (displays crossed
triangle symbol when inO2 alarms are turned
off)
E Abbreviation for "expiratory"
F etO
2 value – the highest level of O2 in the
airway during the expiratory phase within the
measurement interval
G etO
2 alarm limits (displays crossed triangle
symbol when etO
2 alarms are turned off)
NOTE
O2 monitoring is available only with Scio Four Oxi
and Scio Four Oxi plus.
O2_pbox_en.pdf