Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 275
SpO2 and pulse rate with Nellcor OxiMax MCable
Patient preparation for SpO2 monitoring
The following tips provide optimal SpO2 monitoring
results but must never replace hospital-approved
practices or manufacturer’s recommendations.
The accuracy of SpO2 monitoring depends largely
on the strength and quality of the SpO2 signal.
If a finger is used as a monitoring site, remove any
nail polish. Cut the finger nails of the patient, if
necessary, for better sensor placement.
Pulses may be counted erroneously due to the
following conditions:
Placement of a sensor that is too tight
Patient experiences hypotension, severe
vasoconstriction, severe anemia, or
hypothermia
Arterial occlusion proximal to the sensor
Patient is in cardiac arrest or is in shock
Bright light causing erratic measurement or
missing values. Cover the sensor with opaque
material if it is likely to be exposed to direct
bright light.
Significant levels of dysfunctional hemoglobins
(HbCO or MetHb)
Intravascular dyes such as indocyanine green
or methylene blue
Excessive patient movement
Venous pulsations
Placement of a sensor on an extremity with a
blood pressure cuff, arterial catheter, or
intravascular line