SpO2 and Pulse CO-Ox monitoring with Masimo SET MCable
266 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Masimo rainbow SET Pulse CO-Ox parameter setup functions
General Masimo rainbow SET SpO2 setup
functions take place in the Pulse CO-Ox page. To
access this setup page, see page 263. Additional
password-protected functions are available (see
page 269).
Selection Available settings Description
Show parameters
(left button)
– SpHb
1)
(default)
–SpOC
– PVI
–SpCO
– SpMet
Selects the parameter for the
parameter 1 location in the
Pulse CO-Ox parameter field.
The associated parameter label
and value have the largest font.
With an Hb sensor, the default
parameter is SpHb. With a CO-
sensor, the default parameter
for the parameter 1 location in
the parameter field changes
automatically to SpCO.
Changes to the parameter
selection are retained if the
same sensor is disconnected
and then reconnected. The
parameter selection changes to
the default selection, if another
Masimo rainbow SET sensor
type is connected.
1)
If the venous blood source was selected for SpHb Cal, the parameter label changes from SpHb
(arterial blood source) to SpHbv.