Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 265
SpO2 and Pulse CO-Ox monitoring with Masimo SET MCable
Averaging time 2 to 4, 4 to 6, 8 (default), 10, 12, 14, 16 s Determines how quickly the
reported SpO2 value responds
to changes in the patient’s
oxygen saturation.
A longer averaging time
provides a more accurate
result. However, in clinical
situations where rapid
physiological changes have to
be monitored, use a shorter
averaging time.
Color Red, green, blue, yellow, light blue, purple,
orange, white (default).
Determines the color of the
waveforms and parameter
labels and values.
Selection Available settings Description
NOTE
The password-protected alarm setting SpO2
sensor off provides additional SpO2 alarm
configuration.