Mainstream CO2 monitoring
344 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Description Cause CO2 waveform
Elevated waveform
baseline with
corresponding increase in
CO
2 level.
Rebreathing due to one of the
following causes:
Disposable airway adapter is
used although the Cockpit is
configured for the reusable
adapter type
Contaminated airway adapter
(dirty window)
–CO
2 zero drift
Insufficient expiratory time
Faulty expiratory valve
Inadequate inspiratory flow
Malfunction of a CO
2 absorber
system
Partial rebreathing circuits
Change in slope of
ascending limb. Possible
absence of an alveolar
plateau.
Obstruction caused by one of the
following:
Partial obstruction in expiratory
limb of breathing circuit
Foreign matter in upper airway
Partially kinked or occluded
artificial airway
Herniated endotracheal or
tracheostomy tube cuff
Bronchospasm
Elevated baseline, with
pronounced slope on
descending limb
Faulty ventilator circuit valve
Rebreathing (see above)