Mainstream CO2 monitoring
346 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Selection Available settings Description
Zero
(Only available if a
CO
2 device is
connected)
None Zeroes the CO2 sensor if necessary. The
CO
2 sensor stores a new isoelectric point for
CO2 measurements.
Scale 0 to 40 mmHg (default), 0
to 60 mmHg,
0 to 80 mmHg, 0 to
100 mmHg
0 to 5 kPa (default), 0 to
8 kPa, 0 to 12 kPa, 0 to
16 kPa
0 to 5 % (default), 0 to 8 %,
0 to 12 %,
0 to 16 %
Adjusts the size of the CO
2 waveform.
Scale settings apply to both Mainstream and
Microstream.
Atm. pressure 570 to 800 mmHg
760 mmHg (default)
Determines the ambient pressure setting of
the sensor and compensates for pressure
effects. Failure to compensate for pressure
can cause inaccurate measurements.
Gas
compensation
Air (default), N
2O/O2,
O2>50%, HeliOx
Compensates for supplemental oxygen or
N2O or Heliox. Failure to compensate for
supplemental oxygen can cause inaccurate
measurements.
RRc apnea time
[s]
Off (default), 10, 15, 20, 25,
30 s
Specifies the time the M540 waits before
reporting a cessation of breathing as an
apnea event.
RRc apnea time [s]
settings apply to both
Mainstream and Microstream.
RRc apnea
archive
Off, Store (default), Str/Rec,
Record
Determines what happens in response to an
apnea.
RRc apnea archive settings apply to both
Mainstream and Microstream.
Airway adapter Reusable (default),
Disposable
Determines the type of airway adapter used
for CO
2 monitoring.
Compensates for the type of airway adapter
that is being used.
Requires matching the adapter with the
configuration setting at the Cockpit; if the
adapters do not match, the CO
2 value
displayed is compromised.