Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 535
Troubleshooting
!! %0 out of range
high
1)
Parameter value or
value is replaced by
*** depending on
the pressure level
The parameter
value is
above/below the
measurement
range of the
monitor.
Check the NIBP
inflation limits and
adjust them if
necessary (for
example, if the wrong
patient category is
selected).
!! %0 out of range
low
1)
Parameter value or
value is replaced by
*** depending on
the pressure level
None NIBP pneumatic
char. needed
Parameter values
are replaced by ***
NIBP hardware
failure in the
M540.
Contact
DrägerService
®
and
take the M540 out of
service.
! %0 blocked Line
1)
The message Last
measurement
failed! is followed
by the message
Blocked line
Parameter values
are replaced by ***
The inflation rate
is too high during
the inflation cycle
or the time to
evacuate
residual cuff
pressure at the
end of the
deflation cycle is
too short.
Select a different
cuff.
Check the hose
and cuff for
damage.
Restart the
measurement. If
the message does
not clear,
DrägerService
®
.
1)
%0 is a placeholder for the parameter label NIBP.
! %0 cannot
measure
1)
The message Last
measurement
failed! is followed
by the message
Cannot measure
Parameter values
are replaced by ***
The pulse profile
is too poor to
establish a
reliable
measurement
(usually due to
persistent motion
artifact)
Check the patient
and treat if
necessary.
Move the cuff to a
limb with less
movement.
Restart the
measurement. If
the message does
not clear, contact
DrägerService
®
.
Priority Message Parameter field Problem Solution