Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 549
Troubleshooting
!! CO
2 warming up The Microstream MCable is
completing its warm-up cycle.
– Wait for the Microstream
MCable to warm up. During
warm-up, the accuracy is
reduced.
– If the message persists longer
than 15 minutes after the sensor
has warmed up and the ambient
temperature is above 10 °C
(50 °F), contact
yDrägerService
®
.
– You cannot zero the sensor
while this message is displayed
and the ambient temperature is
above 10 °C (50 °F).
– When the ambient temperature
is below 10 °C (50 °F), the
message can display longer
than 15 minutes. In this case, it
is possible to zero the sensor
after the message has been
displayed for at least
10 minutes.
None %0 zeroing in
progress
The Microstream MCable is
being reset to zero.
Informational message – no action
required.
!! etCO
2 > (alarm
limit)
etCO
2 < (alarm
limit)
(except inCO
2)
The parameter value is
above/below the set
upper/lower alarm limits.
– Check the patient and treat, if
necessary.
– Change the alarm limits.
!!! %0 apnea Apnea was detected. – Check the patient and treat, if
necessary.
– Check the placement of sensor.
!! RRc out of range
high
The parameter signal is outside
the measuring range of the
patient monitor.
– Check the patient and treat if
necessary.
– Check the equipment and
replace if necessary.
1)
In the parameter field the parameter value is replaced by ***
Priority Message Cause Remedy