Troubleshooting
552 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
None %0 out of range
high
1)
%0 out of range
low
1)
Parameter value The C.O. is greater
than 20 liters/min
or less than
0.5 liters/min
because of:
a physiological
condition
unstable
baseline
incorrect
injectate
volume,
catheter size,
or computation
constant
faulty catheter,
cable, or
cartridge
Check the patient
and treat if
necessary.
Use cooler
injectate.
Enter the correct
values in the C.O.
page.
Repeat the
measurement. If
message persists,
replace faulty
components.
None %0 Check Injectate
Probe
Parameter value The thermistor is
not connected or
became
disconnected
during a
measurement.
Connect the probe
and repeat the
measurement.
None %0 duplicate device
connected
1)
Parameter value Multiple C.O.
sources are
connected. This
includes CCO
devices connected
via the device
connectivity
option.
Disconnect duplicate
C.O. sources.
None %0 Injectate Too
Cold
1)
Parameter value The injectate
temperature is too
cold during the
measurement
process.
Use an injectate
within the correct
temperature
range of –5 °C to
+30 °C (–23 °F to
+86 °F).
Check equipment
and replace if
necessary.
1)
%0 is a placeholder for the parameter label C.O.
Priority Message Parameter field
Problem Solution