Technical data
590 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
LEDs Green LED (device is connected to AC power).
Yellow LED (briefly during startup)
The battery/charging failure indicator lights up yellow
under the following circumstances:
Battery failure (battery depleted, battery not connect-
ed, battery fault), or
Battery charging error, or battery temperature issue
The Cockpit LEDs on the front light up for a few seconds
when the Infinity P2500 is connected to AC mains after it has
been disconnected for an interval to indicate that the LED is
working properly.
Automatic shutdown Overload limit (the device shuts down)
Power overload: exceed 285 W (at system connector)
Current overload: exceed 14 A (short circuit protected)
Output voltage: exceed 32V
Temperature:
internal temperature exceeds 75°C (167 °F)
battery temperature exceeds 50 °C (122°F) when
battery is in charging status after 1 minute.
Electrical protection Class 1
Inrush current 37 A