GBK51948 RHINO2
Installation Manual Issue 5
6.5.1.8 Reversing Beeper
Parameter Possible Values Default HHP Lite Std Adv
Reversing Beeper
Yes
No
No
Note that the Reversing Beeper parameter does not exist in the Wizard since the Motion
Beeper parameter can be used to switch on the beeper for reversing.
When using the HHP to program, use both the Motion Beeper and the Reversing Beeper
parameters, as set out in the table below.
Motion Beeper Reversing Beeper Wizard Equivalent
No Yes Reverse
Yes Yes None
No No None
Yes No Forward/Reverse
6.5.1.9 Beeper Timing
Parameter Possible Values Default HHP Lite Std Adv
Beeper On Time 100 1500ms 300ms
Beeper Off Time 0 1500ms 700ms
These parameters are only used when Enable Beeper has the value 'Yes'.
Beeper On Time and Beeper Off Time together set
the beeper interval time. During Beeper On Time the
beeper emits a sound. During Beeper Off Time the
beeper is silent.
6.5.1.10 Deep Discharge Beeper
Parameter Possible Values Default HHP Lite Std Adv
Deep Discharge Beeper No / Yes Yes
This parameter is only used when Enable Beeper has the value 'Yes'.
Enables beeping if the battery is drained to the cut-off level set by Battery Cut-Off Voltage.
Note
This parameter must be set to comply with the ISO 7176-14 (2008) “over-discharge protection” requirement.
Page 74 - Programming