H50/H51 Hematology Analyzer User Manual Startup
- 25 -
Chapter 7 Startup
Run Sample Screen
Parameter
Background
WBC
ʇ0.2 x10
9
/L
RBC
ʇ 0.02 x10
12
/L
HGB
ʇ 1 g/L
PLT
ʇ 10 x10
9
/L
Startup cycle is dedicated to daily check the
background values of measured parameters.
It must
be launched every day before QC and then before
patient’s analysis.
Before user login, the system will automatically
finish
background tests. A
fter login, the system
directly enters into Run Sample Screen. Operator
can check the background test results on this screen.
If the test results don’t meet requirements, the
system will hint operator to re
-do
Startup
cycle by the message “STARTUP FAILED –
BACKGROUD CHECK FAILED
”.
operator runs sample test
after a failed
startup, all results will be displayed and printed
with the indication
“Startup failed”
to turn on the analyzer and the
syste
m will automatically do STARTUP
. After
Startup, the system will enter into Login
Screen. After login, the system directly enters
into Run Sample Screen
to display the
STARTUP results.