H50/H51 Hematology Analyzer User Manual Service
- 59 -
Firstly perform a Backflush cycle. I
still unresolved, then perform a Clean cycle. I
problem is still unresolved, lastly p
Cleaning cycle
Check if there are
bubbles in WBC bath during the
analysis cycle.
Check the electrode wires.
Firstly perform a Clean cycle. I
unresolved, then perform a Bleach Cleaning
1) Firstly perform a Backflush cycle. If the problem is
still unresolved, then perform a Clean cycle. I
problem is still unresolved, lastly p
Cleaning cycle
Check if there are
bubbles in RBC bath during the
analysis cycle.
Check the level of bubbles in WBC bath during the
first dilution.
Check if the HGB LED light is on.
1) Check the level bubble flow in the WBC bath
during the run cycle.
Perform a new Start Up cycle.
13.6.3 Mechanical Problems Troubleshooting
Problems Solutions
Diluent leaks around the
needle during sample tests
1) Check the rinsing needle block (presence of clots) and clean
it if necessary
Check the rinsing head.
Fail to turn on the analyzer
1) Check the power supply connection wires.
Check the Power Supply block
test results are bad.
Check whether diluent
container is placed at the same level
with the analyzer and whether the diluent
pinched.
Check the diluent’s remaining volume.
No test results are displayed
Check the screen wires which are connected to CPU board.
Check the electrical connection.