H50/H51 Hematology Analyzer User Manual Test Results
- 38 -
Unstable LED light low measurement during
WBC DIF measurement. This may come from
bubble or unexpected phenomenon during
WBC DIF measurement.
12.3 Morphologic Alarms
Screen 11-1
12.3.1 WBC Histogram Alarm
Detailed information in pop-out window
WBC - L1
WBC - L1 : Small Lymphocytes
Presence of platelet aggregate,
erythroblast or small
lymphocytes at the left of L1.
WBC –
WBC – L5 : Immature Cells
Presence of immature cells at
the right of
Morphologic alarms are related to cells
characterization. The morphologic alarms
have the corresponding codes. The codes
are displayed in the flag box. Press the
codes, an enlarged flag box will pop out
to display
the
detailed messages of the
morphologic a
larms
as the left Screen
shows.
Press SERVICE/SETTINGS/LAB
PARAMETERS/CBC THRESHOLDS
AND FLAGS
and the system will
Screen11-1 as
the left screen shows.
Morphologic alarms
will be affected if the
thresholds displayed in the left screen are
changed. Therefore,
only technical
recommended
to edit values
on this screen.