H50/H51 Hematology Analyzer User Manual Test Results
- 37 -
12.2 System Alerts
Table 11-1
Detailed information in
pop-out window
Triggers
S-
Startup Failed
Background test results don’t meet
background requirements.
INS-T
Instrument temperature out
Means that the Instrument temperature is
lower than 1
tests are not done during the day.
test failed.
RBC
Clog detected during RBC or PLT
measurement (aperture clog).
WBC
Clog detected during WBC measurement
(aperture clog).
WBC-
WBC Balance
Unbalanced counting between the resistive
and the optical measurement on the WBC
Channel. This may be explained by a clog on
the WBC aperture, a clog on the WBC flow
cell or that the stop cock has been left as
closed.
It will trigger an automatic rinsing cycle
called
“APERTURES RINSING”
The system alerts have the corresponding
abbreviations. The abbreviations are
displayed in the flag box. Press the
abbreviations, an enlarged
flag box
will
pop out to display its detailed information
as the left Screen shows.
-1 lists
all the
system alerts and its corresponding
abbreviations.