H50/H51 Hematology Analyzer User Manual Test Results
- 40 -
an absolute limit OR a percentage of the RBC.
NOTE:
Only technical persons are recommended to edit thresholds (CR1 and CR5) and flags in Screen
11-1 because any change might trigger or affect the alarm
RBC - Micro and RBC - Macro.
12.3.3 PLT Histogram Alarm
Morpho Alarms
Detailed information in pop-out window
Triggers
PLT -
Debris
PLT - Debris
Presence of abnormal quantity
of debris or small cells.
PLT -
Schizo
PLT - Schizo
Presence of schizocytes.
PLT - Micro
PLT - Micro
Presence of microcytes.
1) The alarm PLT - Debris will be triggered if this population at the left of CP1 is higher than
an absolute limit OR a percentage of the PLT.
2) The alarm PLT - Schizo will be triggered if this population between CP2 and P is higher than
an absolute limit OR a percentage of the PLT.
3) The alarm PLT - Micro will be triggered if this population defined by CP3-2 width at the
right of CP3 is higher than an absolute limit AND a percentage of the PLT.
NOTE:
Only technical persons are recommended to edit thresholds (P, PL, CP1, CP2, CP3 and CP3-2)
and flags in Screen 11-1 because any change might trigger or affect the alarm
PLT Debris,
PLT - Schizo and PLT - Micro.
fL
Counts