H50/H51 Hematology Analyzer User Manual Test Results
- 41 -
12.3.4 WBC Scattergram Alarm
Screen11-2
Detailed information in pop-out
window
Causes
DIF Noise
– N1: Debris or Small Cells
Presence of platelet aggregate, debris
or resistant RBC.
– N2: Debris or Small Cells
hrolysis
Presence of platelet aggregate,
erythroblast or small lymphocytes.
RL
Review the DIF
Review on Smear
Presence of atypical lymphocytes or
basophiles.
Press SERVICE/SETTINGS/LAB
PARAMETERS/
DIF FLAGS and the
enter into Screen11-2 as
the left
screen shows. The following
m
will be affected if the values
displayed in the left screen are changed.
Therefore,
only technical persons are
to edit values on this screen.