H50/H51 Hematology Analyzer User Manual Test Results
- 41 -
12.3.4 WBC Scattergram Alarm
Screen11-2
Morpho Alarms
Detailed information in pop-out
window
Causes
DIF Noise
Review on Smear:
DIF
N1: Debris or Small Cells
Platelet Aggregates
Presence of platelet aggregate, debris
or resistant RBC.
Review on Smear:
DIF
N2: Debris or Small Cells
Probable Incomplete Eryt
hrolysis
Platelet Aggregates
Erythroblasts
Presence of platelet aggregate,
erythroblast or small lymphocytes.
RL
Review the DIF
Review on Smear
Presence of atypical lymphocytes or
basophiles.
Press SERVICE/SETTINGS/LAB
PARAMETERS/
DIF FLAGS and the
system will
enter into Screen11-2 as
the left
screen shows. The following
m
orphologic
alarms
will be affected if the values
displayed in the left screen are changed.
Therefore,
only technical persons are
recommended
to edit values on this screen.