EasyManua.ls Logo

EDAN H50 - Chapter 9 Calibration; Analyzer Calibration Process; Calibrator Information Management

EDAN H50
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
H50/H51 Hematology Analyzer User Manual Calibration
- 29 -
Chapter 9 Calibration
The H50/H51 Hematology Analyzer has been calibrated prior to delivery. However, errors may
still be introduced to the system during transport or installation. To maintain the accuracy and
precision of this analyzer, Edan recommends that QC test needs to be conducted prior to being
used for sample analysis for the first time and after every repair and maintenance of the analyzer.
If the QC test fails, which is not caused by the improper operation or the controls, the analyzer
needs to be calibrated.
NOTE:
Only the calibration materials provided by EDAN or its authorized distributors should be
used.
Press Calibration to enter into Screen 8-1. Press Change to enter into Screen 8-2. Press Results
to enter into Screen 8-3.
Screen 8-1
Screen 8-2
CHANGE” button allows editing new calibration
information
or re-editing the existent
calibration
information. But all its associated results will be
deleted if you re
-edit the existent calibr
ation
information.
You can directly go to “RESULTS” to do
calibration
test or view calibration
test results. The
sampling needle will move down to the sampling
position automatically. Present calibration
materials
under the sampling needle and press the
start
key
, the analyzer will automatically do the
sample aspiration and analysis, and finally display
the test results on the screen.
The Screen 8-2 is mainly used to edit
calibration
material’s information.
Two ways to edit
Lot, Expiry, Targets
and
Limits
on the Screen 8-2:
Enter manually by Alpha-
Numeric touch
screen keyboard
Import the information from USB drive
with the button “LOAD

Table of Contents

Related product manuals