H50/H51 Hematology Analyzer User Manual Calibration
- 30 -
Screen 8-3
NOTE:
1. 3 calibration tests at least and 12 calibration tests at most are required to be
conducted successively.
2. The column allows selecting or unselecting a result in the table. Mean, SD and
CV are only decided by the selected results ( ) in the table. Automatic calculations
will be implemented accordingly.
3. Concerning calibration materials’ usage, storage, handling and other details, please
strictly respect its user manual.
4. If the room temperature changes more than 10°C during the working day, the
instrument must be calibrated more frequently.
The Screen 8-3 is mainly used to perform
calibration
test and view the test results.
Under the screen, the sampling needle will
move down to aspirate the prepared
calibration
materials by pushing the start
key
and then the system will automatically finish
the whole analysis and display the test results
on the screen.
After tests, press Calibrate and the Screen 8-4
pops out.
Unselect
parameter(s) you do not
want to calibrate.
Press to
finish
calibration. Finally the coefficients will be
updated on Screen 8
-1.
to restart the calibration