H50/H51 Hematology Analyzer User Manual Service
- 50 -
13.5.2 Date/Time Setting
Date/Time Screen
13.5.3 Automatic Cycles Setting
Press Automatic Cycles to enter into Automatic Cycles Screen. You can perform the following
settings on the screen.
Automatic Cycles Screen
Auto Clean: If the operator sets the auto clean frequency (analysis runs number), the system will
launch clean operation at the designated frequency. The auto clean frequency ranges from 10 to
5000 and the default frequency is 100. The operator can change the default value in accordance
with actual needs.
Auto Shutdown: If the operator enables this function, the system will shut down after the
designated idle duration. The idle duration ranges from 30 minutes to 720 minutes and the default
idle duration is 180 minutes. The operator can change the default value in accordance with actual
needs.
NOTE: mn means minute.
Press Date/Time to enter into the Date/Time
Screen.
n this screen, y
ou can select date and time
format and set the current time and date as
the system’s time and date.
Automatic Power up Setting: If you
enable this function, the system will
automatically switch on at your
designated time.