13.7 Maintenance
13.7.1 Automatic Bleach Cleaning Maintenance
Maintenance Screen
Screen 12-13
Screen 12-14
13.7.2 Regular Maintenance
The following maintenance table is recommended to user when:
z 50 samples are tested per day.
z 6 x 52 working days per year
Press Maintenance to enter into Maintenance
Screen.
On the Screen
, the operator can
determine the frequency of performing bleach
cleaning cycle
by setting the following items.
No matter which one
is reached,
the system will
pop out Screen 12
-
13 to prompt users to perform
bl
each cleaning cycle
to turn off analyzer.
Number of Cycles (ranging from 50-
and the default value is 300)
Bleach Frequency (Days)
30 and the default value is 7)
On Screen 12-13, select Bleach + Shutdown
and then press
, the system will
automatically perform bleach cleaning cycle
and finally shut down the analyzer.
If you select Shutdown on Screen 12-13
and then press
to shut down the
analyzer directly. The system will pop out
Screen 12
-
14 when you turn on analyzer
next time to prompt users to do the bleach
cleaning cycle.