H50/H51 Hematology Analyzer User Manual Analytical Performance
- 67 -
A2.5 Accuracy
Parameter
Detection Range
Limit (%)
WBC
3.5×10
9
/ L ~9.5× 10
9
/ L
±6
RBC
3.8× 10
12
/ L ~5.8× 10
12
/ L
±4.0
HGB
110 g/L ~180 g/L
±3.0
PLT
125×10
9
/ L ~350×10
9
/ L
±13.0
MCV
82 fL ~100 fL
±6.0
HCT
35% ~55%
±6.0
A2.6 Measurement Range
Parameter
Units*
Measurement Range
WBC
10
9
/L
0-150
RBC
10
12
/L
0-15
HGB
g/L
0-25
HCT
%
0-100
MCV
fL
0 - 150
PLT
10
9
/L
0 - 4000
RDW-CV
%
0 100
RDW-SD
fL
0 100
MPV
fL
0 50
MCH
pg
0 50
MCHC
g/L
0 50
PCT
%
0 100
PDW
%
0 50
PLCR
%
0 - 100
PLCC
10
9
/L
0 - 4000
LYM, MONO, NEU, EOS,
BASO, ALY, IMM #
10
9
/L 0-150
LYM, MONO, NEU, EOS,
BASO, ALY, IMM %
% 0-100