H50/H51 Hematology Analyzer User Manual Sample Test
- 33 -
Screen 10-3
The operator can edit the following information in Screen 10-3.
1. NAME: Patient name (1-20 characters)
2. PID: Patient identification (1-16 characters)
3. SID: Sample identification(1-16 characters)
4. Birth Date: patient’s birth date
5. Doctor: Doctor ID or doctor’s name (1-16 characters)
6. Comments: Add comments if needed
7. Gender: F(female)/M(male)/U(unknown)
8. Group: Reference Range
9. Mode: DIF WB, DIF PD, CBC WB, CBC PD, DIF CAP, CBC CAP
NOTE:
1) WB stands for whole blood and PD stands for pre-dilution. If PD mode is selected,
the button “DIL DISPENSE” will appear on the Screen 10-1. Press the button and
the system will dispense 300μL dilution via the sampling needle in to the prepared
tube.
2) If the sampling needle is not down in sampling position, press on start key and
needle will go down in sampling position
The Screen 10-3 is used to edit sample’s
information.
After editing the following
information, the system will go back to the
Screen 10
-1 for sample test.