H50/H51 Hematology Analyzer User Manual Safety Guidance
- 2 -
WARNING
9. Connect the analyzer to a grounded socket and make certain that the power supply
meets the requirements specified in the user manual.
10. Do not exceed the maximum permitted load when using multiple portable socket-
outlets to supply the system.
11. SHOCK HAZARD - Do not attempt to connect or disconnect a power cord with wet
hands. Make certain that your hands are clean and dry before touching a power cord.
12. If you have any questions about the power adaptor or the power cord prior to using,
consult EDAN or its authorized distributors for service.
13. The analyzer is not waterproof. Do not use it in locations where water or any liquid
leakage may occur.
14. To avoid the possibility of electrostatic shock and damage to the system, avoid using
aerosol spray cleansers on the analyzer screen.
15. EXPLOSION HAZARD The analyzer is not suitable for use in the presence of a
flammable anesthetic mixture with oxygen or other flammable compounds.
16. To avoid electrical shock, never use the system in altitude exceeding 3,000 meters
above sea level.
17. LASER RADIATION The bar code scanner is a laser product. Do not view beam
generated by the scanner with eyes directly.
18. Periodically have the safety of the analyzer checked by a qualified service engineer.
19. Only accessories supplied or recommended by the manufacturer should be used.
Otherwise, performance and electric shock protection can‘t be guaranteed.
20. Blood samples should be collected according to proper medical guidelines which
contains collection details, such as selection of blood collection site, collection
procedures, sample handling, etc. Sterile techniques should be followed to prevent
the blood collection site from being contaminated.
21. Handle blood samples and collection devices with care, and wear approved
protective gloves to avoid direct contact with samples.
22. To avoid electrical shock and damage to the system, turn off the analyzer and
disconnect the analyzer from the AC power source before cleaning and disinfecting.
23. To avoid being injured, never touch the sampling needle directly with hands.
24. The system is for in vitro diagnostic use only.
25. Perform quality control (QC) tests to make sure that the system works smoothly.
26. The sample collection tube should only be used a single time.
27. The sample is contained in the sample tubes, so sample tubes need to be
disposed of as bio-hazardous waste, complying with local regulatory guidelines.