SCU-1601 Control Unit for Turbomolecular Pump
Table
41 - Error
messages after self test (continued)
1. Connect the cable
correctly. If the cable is not
locked, lock it securely.
2. Set "Leak Valve Option" to
"DISABLE".
1. Check the specifications of
the connected STP pump.
2. Contact Service office.
3. Contact Service office.
4. Contact Service office.
Change the AC input voltage
200 V.
1. Check whether or not a
power failure has
occurred.
2. Connect the power cable
correctly.
3. Replace the power cable.
4. Set the voltage to the rated
voltage
±10%.
1. Disconnection of the
emergency vent valve
cable.
2. Incorrect setting of the
emergency vent
valve.
1. STP pump is not
compatible with external
TMS unit
2. Failure of external TMS
unit
3. Failure of the STP
pump.
4. Failure of the SCU
-1601.
Incorrect connection of the
input power supply
.
1. Power failure.
2. Incorrect connection of the
power cable.
3. Failure of the power cable.
4. Insufficient power supply
voltage.
Touch down. The
magnetic bearing
output stops and the
rotor is descended
on the
touch down
bearing.
Touch down. The
magnetic bearing
output stops and the
rotor is descended
on the
touch down
bearing.
Serial communication
failure.
Disconnection of the
emergency valve cable.
Excessive magnetic
bearing electric current
continues for 30 sec.
Connection of the STP
pump not corresponding to
the input power
supply
(100 V).
Disconnection of the
temperature sensor of the
STP pump.
Insufficient power supply.