STP-iS2207 Series Turbomolecular Pump 
 
MT-89E-001-C
Page 142 
7
SAFETY FUNCTIONS 
 
 
Referred
Section
 
 
 
 
 
The magnetic bearing output stops and the rotor is descended on the touch down bearing.
 
The driver output stops and the rotor continue rotating by inertia. It may take several hours to stop.   
To stop the STP pump quickly, close the vacuum valve at the outlet port flange and introduce gas from the purge port into the STP pump. 
Table 41 - Error List (continued) 
Countermeasures 
 
 
 
  Check whether or not a 
power failure has occurred.
  Connect the power cable 
correctly. 
  Replace the power cable. 
  Set the rated voltage ± 
10 %. 
  Check the volume of water 
and water temperature with 
water cooling unit. In order 
to recover, turn off the 
power (switch "OFF"), and 
then switch "ON" after 
sufficient cooling of the 
control unit. 
(isolate input power supply 
3 minutes or more) 
  Contact Service office. 
Probable causes 
  Failure of the STP pump.
  Failure of the control unit.
isconnection of the motor 
sensor. 
control unit. 
  Power failure. 
 
Incorrect connection of the 
power cable. 
  Failure of the power cable.
  Failure of 
vlotage. 
  Insufficient cooling of the 
control unit. 
  Failure of the control unit.
Occurrence 
condition 
 
motor 
 
 
 control unit. 
Pump operation
*1
   
ecelerate and 
 
 *2
 
ecelerate and 
 
(Error message)
 
 
, 4 
 
"FAILURE"
LED 
Steady red