11BRemote Ports
134/162
Command LC (Last Coupling)
Explanation: query of the current coupling paths. Only with external automatic CDN relevant
Argument:
no argument
Answer: Characters
IMP-OUT, L-N, L-PE, N-PE
Example: LC?
Answer: IMP-OUT
Command VPK (Voltage PeaK)
Explanation: query of the Surge voltage peak measurement [in V] of the last pulse.
Argument:
no argument
Answer: Integer 0..5000
Example: VPK?
Answer: 2345 (positive Impulse)
or Answer: -2100 (negative Impulse)
Command IPK (current PeaK)
Explanation: query of the Surge peak current measurement [in A] of the last pulse.
Argument:
no argument
Answer: Integer 0..2500
Example: IPK?
Answer: 1345 (positive Impulse)
or Answer: -1100 (negative Impulse)
Command ID (IDentification)
Explanation: Inquiry of the type of equipment.
Argument:
no argument
Answer: Characters : TRA v.vv
v.vv stays for the software version
Example: ID?
Answer: TRA 1.15
Command ID (Full IDentification)
Explanation: Inquiry of the type of equipment.
Example:
FID?
Answer: TRA3000 F-S-D-V