4.1.1.5 Measuring outputs EUT Power Voltage (9) and Current (10)
A signal corresponding to the mains voltage is available at these two BNC outputs „EUT power“. Maximum
12 V for the voltage at the output (9) and maximum 12 V for the current at the output (10).
4.1.1.6 Measuring outputs SURGE Voltage (11) and Current (12)
During SURGE tests, voltage sequence of the SURGE waveform can be measured at the output socket
(11) and the current sequence at output socket (12). The range and the accuracy of the measuring system
is given in the Chapter 1.2 Technical data Section 1.2.8 measuring circuits, measuring outputs.
4.1.1.7 Trigger output for oscilloscope (13)
This output provides all the necessary trigger impulses for the different tests. The different trigger levels
and the time delays are listed in Chapter 1.2 Technical data paragraph 1.3.13
4.1.1.8 The Push-button ENTER (14)
Numeric values are changed only when the ENTER button is pressed.
4.1.1.9 Push-button Edit (15)
This button has a multifunctional use:
• Activate the dialogue line
• Open pull down windows
4.1.1.10 Buttons F1 to F6 (17)
The buttons F1 to F6 are allocated to the function as indicated on the display. Depending on the menu,
different functions are allocated to the six buttons.
4.1.1.11 Numeric control panel (18)
If the cursor is activated in one line of the display, then data can be input with the numeric key board. Each
data input must be terminated with ENTER.
The button BSP (Backspace) enables correction of an incorrect data input.
4.1.1.12 Dialogue line within the display (5)
Indicates what range can be selected or which next step must be done.
4.1.1.13 USB button (5)
All service date is stored to the USB stick when inserted into the rear panel “USB Port”