4.2.1.9 EUT Power 2 Inputs (48)
Input for the disturbance level during interruption. When an external source, e.g. external variac or an
external ac/dc source (PS3), is used, the external source must be connected to these inputs (48).
Supply data a.c.: 20 to 250 V 16 A. no derating necessary
Supply date d.c.: 20 to 300 V 10 A. see derating below. Only applicable for TRA versions with TRA3000
..D-V included.
A temperature sensor protects the high voltage switch in case of d.c. supply.
4.2.1.10 Interface „Port 1“ RS232 for controller PC (49)
The RS232 interface port can be used to control the TRA3000 F-S-D-V-C using an external PC with
customized program. To configure the interface, see Chapter 13 „Remote Control“.
4.2.1.11 Interface „Port 2“ RS 485 for controlling external coupling networks or checking the EUT
failed status (50)
Via this interface, the coupling path of external CDNs can be controlled.
For further information, see the specific CDN manual.
4.2.1.12 The USB Port (51)
Via this interface, the service data can be stored on an USB stick. In case of failure the data can be sent via
Email to EMC PARTNER for analysis.
4.2.1.13 Forced cooling of the TRA3000 F-S-D-V-C (56)
A ventilator cools the TRA3000 F-S-D-V-C internally. Forced cooling is necessary for the impulse forming
network devices and the electronic high-voltage switch. A distance of about 20 cm must be maintained
between the rear panel of the TRANSIENT 3000 and any wall, and about 3 cm between the sides of the
TRA3000 F-S-D-V-C and any equipment or wall. The TRA3000 F-S-D-V-C can be built into a 19“ rack, with
3 cm side separation.
4.2.1.14 Ethernet remote control (57)
The Ethernet interface port can be used to control the TRA3000 F-S-D-V-C using an external PC with
GENECS-TRA program. To configure the interface, see Chapter 13 „Remote Control“.