TRANSIENT3000
49/162
4.2.1.15 Emergency stop, ( EMERGENCY STOP ) (58)
When the „emergency stop“ input is activated, the EMC test and the EUT power supply will be immediately
interrupted. The power supply of the TRA3000 F-S-D-V-C will not be turned off. The status „emergency
stop“- will be indicated on the front panel. Emergency stop corresponds to 0V at the input.
4.2.1.16 EUT Failed input (59)
This input can be used for a single channel EUT passed/failed detection during the EMC test. EUT failed is
equal to 0V.
10R
EUT-FAILED
BNC
CTRL3000
10K
+24V
27V
The trigger value at the input is approximately 5Vdc (active low). The
maximum input value is 24Vdc.
Driving with an open-collector output is recommended.
10R
EMERGENCY
STOP
BNC
CTRL3000
10K
+24V
27V
The trigger value at the input is approximately 5Vdc (active low). The
maximum input value is 24Vdc.
Driving with an open-collector output is recommended.